Technical data
SAPHIR was designed to meet the following criteria:
- chemically inert internal surfaces
- large volume to surface ratio
- clean air supply for simulation of remote atmospheres
- variable access to the measurement volume for instruments
- international platform for basic research and for intercomparisons of measurements
- wide range of standard equipment for trace gas measurements
Parameters | Instrument |
---|---|
hydroxyl-radicals (OH) | laser absorption spectroscopy (DOAS) |
hydroxyl-radicals (OH), and HO2, RO2 | laser-induced fluorescence (LIF) |
OH reactivity | Laser Photolysis - Laser Induced Fluorescence (LP-LIF) |
hydrocarbons up to C6 | gas chromatography |
volatile hydrocarbons up to C10 | gas chromatography-mass spectrometry (GC-MS) |
carbon monoxide (CO) | gas chromatography |
PAN (peroxyacetyl nitrate) | gas chromatography |
hydrocarbons | proton-transfer-reaction time-of-flight mass spectrometry (PTR-TOF-MS) |
ozone | UV absorption-spectrometry |
solar radiation | spectroradiometer |
nitric oxide, nitrogen oxides | chemiluminescence |
NO3 / N2O5 | Cavity ring-down spectroscopy |
CO, CO2, H2O, CH4 | Cavity ring-down spectroscopy |
aerosol properties | CPC, SMPS |
aerosol composition | aerosol mass spectrometry (AMS) |
temperature, humidity, air pressure | standard instruments |
Detailed technical data: | |
---|---|
shape: | cylinder |
volume: | 270 m3 |
surface: | 324 m2 |
surface/volume ratio: | 1.08 m-1 |
diameter: | 5 m |
length: | 20 m |
wall material: | FEP-Teflon, double-wall |
wall thickness: | 150 mm |
lighting: | sun |
shading: | louvre system, several stages |
closing time: | 60 s |
gas purity: | N2/O2 6.0 |
maximum purge rate: | 300 m3/h |
temperature range: | outside temperatur |
operation pressure: | 1000 hPa |
Last Modified: 25.03.2025