Quartz Crystal Microbalance

Quartz Crystal Microbalance
Quartz crystal microbalance with dissipation monitoring system

Quartz crystal microbalance with dissipation monitoring system (QCM-D) measures change of mass and viscoelasity of material on the surface in real time. Frequency and dissipation are monitored simultaneously and give information about mass and rigidity. Adsorption/desorption, chemical reaction, and structural change on the surface in liquid can be analyzed in nanogram sensitivity.

Applications:

  • Biomaterials
  • Energy
  • Environmental
  • Cell & Molecular Biology
  • Biomolecules (DNA, Lipids, and Proteins), surfactants, nanoparticles and polymers

Options:

  • Q-sense E1/E4
  • Flow module
  • Window module
  • Open module
  • Electrochemistry module

Specifications:

  • Number of sensors 1(E1), 4(E4)
  • Volume above each sensor ~ 40μl
  • Minimum sample volume ~ 300μl
  • Working temperature 18 to 45 °C, controlled via the software, stability ± 0.02 K
  • Cleaning All parts exposed to liquid can be removed and cleaned in e.g. ultrasound bath
  • Sensor crystals 5 MHz, 14 mm diameter, polished, AT-cut, gold electrodes
  • Frequency range 1-70 MHz (up to the 13th overtone, 65 MHz for a 5 MHz crystal)
  • Maximum time resolution, 1 frequency up to 200 data points per second
  • Maximum mass sensitivity in liquid ~ 0.5 ng/cm2(5pg/mm2)
  • Normal mass sensitivity in liquid ~ 1.8 ng/cm2(18pg/mm2)
  • Maximum dissipation sensitivity in liquid ~ 0.04 x 10-6
  • Normal dissipation sensitivity in liquid ~ 0.1 x 10-6
  • Typical noise peak to peak (RMS) in liquid ~ 0.16 Hz (0.04 Hz)

More information

Last Modified: 31.05.2022