SAXS

GANESHA

SAXS and WAXS Laboratory Beamline equipped with a Ga-Metal-jet X-ray source and EIGER 4M single photon counting detector. The main focus of the instrument is to be very versatile and flexible in terms of sample environments and covered q-ranges. Having a Metal-jet source with its high intensity and brilliance enables us to measure all samples in air. The evacuated tube holding the movable detector can be shifted along the beam path to vary the available sample area from a few millimeters up to ∼1m.

SAXS/WAXS with EIGER 4M

0.03 – 22nm-1

WAXS with PILATUS100K (off-center)

∼10-70nm-1

Standard sample stages:

Capillary holder for up to ten 2mm capillaries for liquid samples from -5 to 80°C (fully integrated)
Linkam HT600 single capillary 1-1.5mm or flatcell with mica window -10 to 400°C (fully integrated)
Linkam TST350 Tensile stage with force measurement -10 to 350°C also with mica windows
Linkam CSS450 shear device with kapton or scotchtape windows
GISAXS with sample tilt on extra stage with 5 motors (x,y,tilt-x,tilt-y,rotation-z)
Various scaffolds for solid samples to measure with minimal background

Special setups:

5 motor stage with needle tip for rotating crystal-XRD for colloidal crystals
Magnetic sample holders for capillaries with up to ∼0.7T
Flow capillary with peristaltic pump
Microfluidic chips with fitted thin capillaries (0.5 to 0.1mm) and syringe pumps
Nanoparticle kinetics in specially made X-Ray transparent reaction vessel with up to 380°C

Details.pdf

SAXSPACE

SAXSPACE

Kratky Camera (Anton Paar)

SAXS

nm-1 bis 6 nm-10.1

WAXS

0,5 nm-1 bis 18 nm-1

Temperature

-10°C -300°C, auto sampler

CCD camera detector

 

Photon flux at the sample 1.3x107 (point-), 1x109 (line collimation)

 

Typical use case: protein solutions, polymer solutions, SEC-SAXS

Further information: Link pdf

NANOSTAR U: (Bruker AXS)

The applications range from polymers to biology or liquids and are limited to the scattering vector range given by the wavelength of the xray. Powerful combinations arise if both neutron and xray scattering are applied to the same system, exploiting the full contrast program. The scattered beam contains 3dimensional information on the spatial structure.

Rotating Anode

Cu Ka

Cross-coupled Göbel Mirrors

 

Detector distances

230 < D < 1060 mm

Scattering vector q

0.006 < q < 3 A-1

Detector

2d VANTEC 2048x2048 resolution

Temperature control

-30°C < T < +300°C

In-situ tensile testing

Up to 400% strain, 0.001 < v < 1mm/s

DSC_0482_300px_jpg.poster.jpeg

Contact:

Dr. Martin Dulle

Phone: +49 2461 61-85160
Email: m.dulle@fz-juelich.de

Dr. Ralf Biehl

Phone: +49 2461 61-4685
Email: ra.biehl@fz-juelich.de

ORCID